GET STARTED

0$

Redirection to the technical assistance portal

XLASE PLUS

Troubleshooting manageable by the client

Field Required to Continue

Next Step

1.1 The device does not turn on

The touchscreen display does not light up, it remains disactivated.


Make sure the cable is properly plugged into the machine and into the power outlet.

Field Required to Continue

Next Step

3.1 Inside the cooling circuit there are air bubbles

Inside the cooling circuit there are air bubbles that do not allow the liquid to circulate properly.


Perform Fill and Vent procedure.

Field Required to Continue

Next Step

2.1 The touchscreen display appears to be on

There is no response when typing or selecting commands.


Turn the device off and wait for two minutes. Then turn it back on.

Field Required to Continue

Next Step

3.2 Not enough liquid

There is not enough liquid inside the cooling circuit.


Perform Fill and Vent procedure.

Field Required to Continue

Next Step

1.2 The device does not turn on

The touchscreen display does not light up, it remains disactivated.


Make sure the cable is working properly by connecting another electronic device.

Field Required to Continue

Next Step

1.3 The device does not turn on

The touchscreen display does not light up, it remains disactivated.


Replace the cable with a similar one, if available.

Field Required to Continue

Next Step

1.4 The device does not turn on

The touchscreen display does not light up, it remains disactivated.


Check the integrity of the fuses with the machine disconnected from the power supply. Open the special compartment located near the power socket. Carefully check that the fuse is not blown or that the wire inside is not broken.

Field Required to Continue

Next Step

2.2 The touchscreen display appears to be on

There is no response when typing or selecting commands.


Make sure there is no external pressure on the display monitor.

Field Required to Continue

Next Step

4.1 Handpiece recognition error


Check correct connection of the handpiece to the device.

Field Required to Continue

Next Step

Flow Alarm

Troubleshooting manageable by the client

Field Required to Continue

Next Step

Diode handpiece 1200/2800/4000

Troubleshooting manageable by the client

Field Required to Continue

Next Step

4.2 Optical alarm


Check optical alarm and insert it.

Field Required to Continue

Next Step

5.1 While applying pressure to the button / pedal, the handpiece does not emit a flash


Make sure that the device is in operating mode.

Field Required to Continue

Next Step

5.2 Check operation


If the handpiece button does not respond, check operation using the pedal and viceversa.

Field Required to Continue

Next Step

5.3 Pulse counter


Check if the pulse counter on the monitor progresses when the button is pressed. In the event that the pulses which are not emitted are also counted, a power generator failure is assumed.

Field Required to Continue

Next Step

6.2 Check the hydraulic system


Check the hydraulic system. The presence of air in the system could affect the efficiency of the cooling system. Carry out the Fill and Vent procedure.

Field Required to Continue

Next Step

6.1 The diode repeatedly goes into OVEARHEAT


Check temperature. It must not exceed 24 ° C.

Field Required to Continue

Next Step

6.3 Reduce frequency


It is possible, especially in prolonged sessions, that there is no need to reduce the pulse emission frequency. Reduce from display to 5Hz.

Field Required to Continue

Next Step

7.1 Optics efficiency. Sapphire is not perfectly transparent.


An initial condensation that could condense the sapphire tip is a normal condition. However, the emission window must be cleaned constantly, even during treatment, using a soft gauze soaked in an alcohol and water solution.

Field Required to Continue

Next Step

8.1 Presumed reduction of energy in the emission phase


Proceed with the energy measurement if the system allows it

Field Required to Continue

Next Step

CPL Handpiece

Risoluzione Problemi gestibili dal cliente

Field Required to Continue

Next Step

9.1 Handpiece recognition error


Check the correct connection of the handpiece to the device

Field Required to Continue

Next Step

10.1 While applying pressure to the button / pedal, the handpiece does not emit a flash


Verify that the device is in operating mode.

Field Required to Continue

Next Step

10.2 Check operation


If the handpiece button does not respond, check operation using the pedal and viciversa.

Field Required to Continue

Next Step

10.3 Pulse counter


Check if the pulse counter on the monitor progresses when the button is pressed. In the event that the pulses which are not emitted are also counted, a power generator failure is assumed. (open a ticket).

Field Required to Continue

Next Step

11.1 In operating mode the lamp does not turn on (FLASH LAMP ALARM)


Check that at START the Cut Off filter is not inserted in the slot of the handpiece. Switch off the machine and repeat the ignition mode with the cut off deactivated. Activate it only when specifically indicated on the software.

Field Required to Continue

Next Step

12.1 Presumed reduction of energy in the emission phase


Check the cleanliness of the cut off filter. The filter must be perfectly transparent. Clean with a soft gauze soaked in a solution of alcohol and water.

Field Required to Continue

Next Step

12.2 Check the transparency


Check the transparency of the tip of the guide glass.

Field Required to Continue

Next Step

12.3 Energy measurement


Take an energy measurement if the device allows it.

Field Required to Continue

Next Step

12.4 Check that the handpiece is correctly inserted


Field Required to Continue

Next Step

QS ND: YAG Handpiece

Troubleshooting manageable by the client

Field Required to Continue

Next Step

13.1 The lamp does not turn on


Disconnect the handpiece from the connector and insert another if available. In the event that the handpiece used as a replacement should light up, it means there is a problem with the lamps. Therefore open a ticket.

Field Required to Continue

Next Step

14.1 Handpiece recognition error


Check the correct connection of the handpiece to the device

Field Required to Continue

Next Step

15.1 The lamp does not turn on


Disconnect the handpiece from the connector and insert another if available. In case of operation of the latter, the problem is related to the lamp. Open a ticket.

Field Required to Continue

Next Step

16.1 Presumed reduction of energy in the emission phase


Check cleanliness of the anti-reflective lens positioned at the tip inside the handpiece. Clean with a soft gauze soaked in a solution of alcohol and water.

Field Required to Continue

Next Step

16.2 Check that the focal lens


Check that the focal lens inserted in the tip of the handpiece is perfectly clean. Clean with a soft gauze soaked in a solution of alcohol and water.

Field Required to Continue

Next Step

16.3 Energy measurement


Take an energy measurement if the device allows it.

Field Required to Continue

Next Step

17.1 KTP - Presumed reduction of energy in the emission phase


Field Required to Continue

Next Step

17.2 Check that the optics are perfectly clean.


Check that the optics are perfectly clean. Clean with a soft gauze soaked in a solution of alcohol and water

Field Required to Continue

Next Step

17.3 Check for correct alignment of the optics.


Check for correct alignment of the optics. Make sure that the indication on the front of the handpiece matches that on the KPT bushing.

Field Required to Continue

Next Step

SLP ND:YAG Handpiece

Troubleshooting manageable by the client

Field Required to Continue

Next Step

18.1 Handpiece recognition error


Check the correct connection of the handpiece to the device

Field Required to Continue

Next Step

19.1 The lamp does not turn on


Disconnect the handpiece from the connector and insert another if available. In case of operation of the latter, the problem is related to the lamp. Open a ticket.

Field Required to Continue

Next Step

20.1 Presumed reduction of energy in the emission phase


Check cleanliness of the anti-reflective lens positioned at the tip inside the handpiece. Clean with a soft gauze soaked in a solution of alcohol and water.

Field Required to Continue

Next Step

20.2 Check that the focal lens


Check that the focal lens inserted in the tip of the handpiece is perfectly clean. Clean with a soft gauze soaked in a solution of alcohol and water.

Field Required to Continue

Next Step

20.3 Energy measurement


Take an energy measurement if the device allows it.

Field Required to Continue

Next Step

21.1 KTP - Presumed reduction of energy in the emission phase


Field Required to Continue

Next Step

21.2 Check that the optics are perfectly clean.


Check that the optics are perfectly clean. Clean with a soft gauze soaked in a solution of alcohol and water

Field Required to Continue

Next Step

21.3 Check for correct alignment of the optics.


Check for correct alignment of the optics. Make sure that the indication on the front of the handpiece matches that on the KPT bushing.

Field Required to Continue

Next Step

SPACERS


Check the correspondence of the magnets with those on the bushing of the handpiece. Rotate to check for proper anchoring.

Field Required to Continue

Next Step

Check the presence of the magnets


Check the presence of the magnets on the spacer side and the handpiece bushing side.

Field Required to Continue

Next Step