The touchscreen display does not light up, it remains disactivated.
Check the integrity of the fuses with the machine disconnected from the power supply. Open the special compartment located near the power socket. Carefully check that the fuse is not blown or that the wire inside is not broken.
5.1 The handpiece does not create thermal energy on the skin
Check the application mode: in ablaze mode the adhesive electrode must be connected to the patient's back and the handpiece will act as a capacitive electrode. In the endodhape mode make sure the electrodes for closing the electric arc are positioned on the tip of the handpiece. In this case the electrode on the body is not required.
7.1 The handpiece does not create thermal energy on the skin.
Check the application mode: in ablaze mode the adhesive electrode must be connected on the patient's back and the handpiece will act as capacitive electrode.
6.5 Check for the presence of product inside the suction ducts.
Check for the presence of product inside the suction ducts.
In case of residues, remove them from the fluid containers.
Then proceed with cleaning the air ducts.
16.4 Check for the presence of product inside the suction ducts.
Check for the presence of product inside the suction ducts.
In case of residues, remove them from the fluid containers.
Then proceed with cleaning the air ducts.