6.4 Check for any presence of product inside the intake ducts.


Check for any presence of product inside the intake ducts. In case of residues, remove them from the fluid containers. Then proceed with cleaning the air ducts.

Field Required to Continue

Next Step

7.1 In the operative phase, the currents are not felt by the patient


Check that the applicator is correctly positioned in the treatment area, and that the fold of fat is correctly inserted into the cavity

Field Required to Continue

Next Step

3.3 Verify that applicator no. 1 is correctly inserted


Verify that applicator no. 1 is correctly inserted in the multi-connection socket and that the clamping bush is fully screwed up to the stop

Field Required to Continue

Next Step

4.2 Check the temperature


Field Required to Continue

Next Step

4.3 Check that there is at least 70 cm of free area around the device


Field Required to Continue

Next Step

5.1 In operating mode, the applicator does not reach the desired cooling temperature within 5-10 minutes.


Field Required to Continue

Next Step

CRYOLIPOSCULPT

Troubleshooting manageable by the client

Field Required to Continue

Next Step

1.1 The device does not turn on

The touchscreen display does not light up, it remains disactivated.


Make sure the cable is properly plugged into the machine and into the power outlet.

Field Required to Continue

Next Step

3.1 Inside the cooling circuit there are air bubbles that do not allow the liquid to circulate properly.


Field Required to Continue

Next Step

2.1 The touchscreen display appears to be on

There is no response when typing or selecting commands.


Turn the device off and wait for two minutes. Then turn it back on.

Field Required to Continue

Next Step

3.2 There is not enough liquid inside the cooling circuit.


Field Required to Continue

Next Step

1.2 The device does not turn on.

The touchscreen display does not light up, it remains disactivated.


Make sure the cable is working properly by connecting another electronic device.

Field Required to Continue

Next Step

1.3 The device does not turn on

The touchscreen display does not light up, it remains disactivated.


Field Required to Continue

Next Step

1.4 The device does not turn on

The touchscreen display does not light up, it remains disactivated.


Check the integrity of the fuses with the machine disconnected from the power supply. Open the special compartment located near the power socket. Carefully check that the fuse is not blown or that the wire inside is not broken.

Field Required to Continue

Next Step

2.2 The touchscreen display appears to be on

There is no response when typing or selecting commands.


Make sure there is no external pressure on the display monitor.

Field Required to Continue

Next Step

4.1 The system shows the message "COOLING PHASE WAIT" and the emission of the applicators stops


Field Required to Continue

Next Step

Flow Alarm

Troubleshooting manageable by the client

Field Required to Continue

Next Step

6.1 There is insufficient suction inside the Cryoliposculpt applicator.


Check that the wool filter is new and correctly positioned inside the right hole

Field Required to Continue

Next Step

6.2 Check the liquid container


Check that the liquid container on the back of the machine is correctly screwed

Field Required to Continue

Next Step

6.3 Check the O-ring


Check that the sealing O-ring is present on the cup

Field Required to Continue

Next Step