The touchscreen display does not light up, it remains disactivated.
Check the integrity of the fuses with the machine disconnected from the power supply. Open the special compartment located near the power socket. Carefully check that the fuse is not blown or that the wire inside is not broken.
5.1 The handpiece does not create thermal energy on the skin
Check the application mode: in ablaze mode the adhesive electrode must be connected on the patient's back and the handpiece will act as a capacitive electrode. In endodhape mode, check that the electrodes for closing the electric arc are positioned on the tip of the handpiece. Here the electrode on the body is not required.
7.1 The handpiece does not create thermal energy on the skin
Check the application mode: in ablaze mode the adhesive electrode must be connected on the patient's back and the handpiece will act as a capacitive electrode.
6.5 Check for the presence of product inside the suction ducts.
Check for the presence of product inside the suction ducts.
In case of residues, remove them from the fluid containers.
Then proceed with cleaning the air ducts.
16.4 Check for the presence of product inside the suction ducts
Check for the presence of product inside the suction ducts.
In case of residues, remove them from the fluid containers.
Then proceed with cleaning the air ducts.
Enter your personal details and you will be redirected to the technical assistance portal, where you can open a ticket and be contacted by a Biotec Italia specialist technician
You will be redirected to your personal account Biotec Italia